HL-RC-T-US

ขอใบเสนอราคา

 

OVERVIEW

Wireless temperature logger 915 MHz (US version) with battery monitoring.

Both the ROTRONIC HW4 software and one of the available RF adapters are required to program the data logger, and to retrieve / analyze the data. HW4 can also be used to display and record data in real time on a PC or laptop. Each RF adapter can connect up to 100 loggers (wireless RF network). Both the HW4 software and data loggers are FDA 21 CFR Part 11 and GAMP compatible. A validated version of the HW4 software is available.

Applications
Phamaceutical and food industries, meteorology, environmental engineering, museums/glass cabinets, monitoring of storerooms, mechanical engineering, chemical industry, research and development

Features

  • Accuracy: ± 0.1 K at 0 °C
  • Temperature operating range: -40…+85 °C
  • Integrated battery monitoring (HL-RC only)
  • High storage capacity: up to 300,000 measured values with serial number, time and date
  • Flash memory for data security in the event of power failures
  • Long-term recording up to 6 years without battery replacement
  • Transmission distance: up to 300 m (free field) with remote probe, 100 m with integrated temperature probe
  • Data security: PIN for activation and data access
  • Plastic housing, white, IP69 (submersible)

TECHNICAL DATA

PROBE TYPEPt-1000
OPERATING RANGE-40…85 °C / 0…100 %rh
ACCURACY AT 23 ±5 °C± 0.1 K at 0 °C
MEASURING RANGE WITH INTEGRATED PROBE-40…85 °C / 0…100 %rh
POWER SUPPLY / POWER CONSUMPTIONLithium (Li-SOCl2) battery 2400 mAh
TEMPERATURE SENSORPT1000 (metal point at the end of case)
MEMORY SIZE2,048 KB of approximately 500,000 readings, non-volatile, even during a power outage
EXPLOITATION SOFTWAREROTRONIC HW4
DATA LOGGING300’000 measured values with serial number, time and date
USER INFORMATIONfrom Software HW4 (devicemanager)
DATA PROCESSING VIA HW4Yes
DEVICE PROTECTION BY PASSWORDYes
CHOICE OF BATTERY / RECHARGEABLE BATTERYLithium (Li-SOCl2) Battery 2400 mAh (Exchangeable)
COMMUNICATION INTERFACESLAN interface / USB dongle / GPRS logger
ENCLOSURE MATERIALPOM
IP PROTECTION RATINGIP69 (submersible)
FDA/GAMP COMPATIBILITYFDA 21 CFR Part 11 and GAMP compatible
DIMENSIONS140 x Ø 20 mm
WEIGHT80 g

DOWNLOADS

Wireless Datalogger System – CE declaration (63.32 KB)11/10/2015

Wireless Datalogger System – Data sheet (610.52 KB)11/6/2015

Wireless Datalogger System – Manual v1.02 (802.04 KB)3/15/2016