Contact Us: | contact@sampledomain.com

Accumac / Probe /AM1610 Precision Industrial PRT

AM1610 Precision Industrial PRT

· Sensor Length 5 mm

· Accuracy ±0.05°C

· Durable and shock resistance

· Customized dimensions available


With 5mm long sensor, AM1610 series precision industrial PRT provides users an excellence temperature probe with extremely low stem conduction (heat loss along sheath) in precision temperature measurement. It is a great tool for measuring and testing the vertical temperature uniformity of heat sources.

AM1610 uses a special aged and tested 5-mm long sensing element to achieve the best performance in stability and repeatability. A unique assembly procedure provides the best balance among the hysteresis effect, mechanical shock and thermal shock performance.

Features

· Temperature range: -60 °C to 160 °C

· Sensor Length: 5 mm

· Accuracy: ±0.05 °C

· Long term drift: ±0.04 °C

· Short term stability: 0.01 °C

· Durable and shock resistance

· Temperature Coefficient 0.00385

· SST sheath

· Quick response time

· Customized dimensions available

Temperature Range -60°C to 160°C
Resistance at 0 °C Nominal 100 Ω
Temperature Coefficient 0.00385 Ω/ Ω/°C
Accuracy ±0.05°C
Drift ±0.04°C at 0 °C after 100 hours at 160 °C
Short Term Stability ±0.01°C
Thermal Shock ±0.01°C after 10 times thermal cycles from minimum to maximum temperatures
Hysteresis <=0.01°C
Self-heating 75 mW/°C
Response Time 4 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current 0.5 mA or 1 mA
Sensor Location 5 mm
Sensor Location 3 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Stainless Steel 316L
Dimension 1610-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1610-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)
External Leads Teflontm –insulated copper wire, 4 leads, 2.5 meters
Handle Dimension 1610-12: 15mm (OD) X 65 mm (L)
1610-9: 10mm (OD) X 50 mm (L)
Handle Temperature Range* -30°C to 150°C
Calibration Options 1610-12-T, PRT with NIST traceable calibration and data
1610-9-T, PRT with NIST traceable calibration and data

Model Description
9001 Carrying Case