5080A High Compliance Multi-Product Calibrator

ขอใบเสนอราคา

 

 

Description

Accurate, reliable analog meter calibration
The 5080A Multi-Product Calibrator calibrates your analog workload accurately and reliably, thanks to its high voltage and current compliance. With maximum burden up to 800 mA for ac/dc voltage, and voltage up to 50 V for ac/dc current, 5080A calibrators can drive a wide range of analog meters.

Options and accessories expand workload coverage
Options and accessories enable you to use the 5080A Multi-Product Calibrator to calibrate an even broader workload, including:
Clamp meters. The 9100-200 10/50 turn coil and 5500A/COIL 50-turn current coil enables the 5080A to calibrate most popular clamp meters at currents up to 1000 A rms amps.

  • Oscilloscopes. Oscilloscope calibration to 200 MHz quickly, easily, and cost effectively. Verify dynamic response, bandwidth, timing, multiple triggering functions, and input resistance.
  • Megohm meters. This option sources high ohms, high voltage resistors up to 18 G ohms. It also measures high voltage outputs.

Protective circuits prevent damage
Mains voltage inadvertently applied to a calibrator’s output terminals can cause extensive damage, requiring costly repairs. Electrical protection for calibrator outputs is vital for daily operation. The 5080A calibrator’s innovative protection circuitry prevents it from being damaged by reversed input voltage, so you can use it with confidence day after day.

Highest voltage and current compliance

The 5080A has the highest voltage and current compliance of any calibrator in the Fluke Calibration multi-product and multifunction calibrator families, making it an ideal solution for calibrating analog meters and other instruments requiring higher drive capability for proper operation.

Maximum burden or compliance voltage
Functions
Model DC voltage AC voltage DC current AC current*
5080A 600 mA 800 mA 50 V 44 V
9100 20 mA 20 mA 4 V 4 V
5500A 10 mA 10 mA 4.5 V 3 V
5520A 10 mA 10 mA 7 V 5 V
5700A 50 mA 50 mA 10 V 7 V
5720A 50 mA 50 mA 10 V 7 V

* With AC LCOMP on

Automate the 5080A Multi-Product Calibrator to increase efficiency and throughput
Versatile software applications enable automated calibration as well as paperless data collection and reporting

  • 5080/CAL. The 5080/CAL software is designed for calibrating analog and digital workload with the 5080A calibrator. It enables you to automate calibration, manage inventory, collect data and print reports, easily and economically.
  • MET/CAL® Lite for 5080A.MET/CAL Lite provides the power of MET/CAL Plus software in a lower cost version designed for use with 5080A calibrators.
  • MET/CAL® Plus. For the full spectrum of calibration automation and asset management, choose MET/CAL Plus software. Add Manual MET/CAL software for mechanical or dimensional workload where automation is not required.The solutions you need, from the leader in calibration

The solutions you need, from the leader in calibration
Fluke Calibration is a premium brand, well known around the world for its accurate, dependable, high quality products. Long known as a leader in dc and low frequency ac calibration, Fluke Calibration is also recognized for its offerings in temperature, pressure, power, process, and rf calibration. Fluke Calibration provides the calibrators, standards, software, service, support and training you need for a complete solution in your cal lab.

Highlights

Calibration solutions for your analog and digital workload

  • High compliance
  • Protection circuitry
  • Calibrates a wide workload, including analog meters and 3.5 and 4.5 digits DMMs
  • Optional 5080/CAL software for easy-to-use, automated calibration

The 5080A Multi-Product Calibrator calibrates your analog and digital workload accurately and economically. Its high voltage and current compliance makes analog workload calibration easy and precise. And built-in protection circuitry protects it against damaging input voltages.
This easy-to-use multifunction calibrator can handle a wide meter calibration workload that includes:

  • Analog meters
  • Panel meters
  • Digital multimeters
  • Watt meters
  • Clamp meters (with coil accessory)
  • Megohm meters (optional)
  • Oscilloscopes to 200 MHz (optional)

Versatile software applications enable you to record paperless results, and more. Best of all, the 5080A Multi-Product Calibrator offers this versatile functionality at a price that fits nicely into your budget.

video

5080A Multi-Product Calibrator Video Brochure

5080A Multi-Product Calibrator Video Demo

Specifications

General Specifications
Warm up Time – Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time – Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces – RS-232 and Ethernet
Temperature Performance – Operating: 0°C to 50°C
– Calibration (tcal):15°C to 35°C
– Storage:-20°C to +70°C
Temperature Coefficient – Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C.
Relative Humidity – Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
– Storage:<95%, non-condensing
Altitude – Operating: 2,000m (6,500 ft) maximum
– Non-operating: 12,200m (40,000 ft) maximum
Safety – Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004
Analog Low Isolation 20V
EMC – Meets EN613261:2006.
Line Power – Line Voltage (selectable): 100V, 120V, 220V, 240V
– Line Frequency: 47 to 63 Hz
– Line Voltage Variation: + 10% about line voltage setting
Power Consumption 600 VA
Calibration Documentation – Report of calibration with data included; 17025 accredited report optional
Dimensions – Height:19.3 cm (7.6 in)
– Width:43.2 cm (17 in), 44.3 cm (17.5 in) including handles
– Depth:53.8 cm (21.2 in)
Weight (without options) 22 kg (48 lbs.)
Absolute Uncertainty Definition – The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated.
Specification Confidence Interval 99%

Models and Accessories

Model Name Description
5080A –  Multi-Product Calibrator
5080A/MEG –  Multi-Product calibrator with Megohm Meter Calibration Option
5080A/SC –  Multi-Product Calibrator with Oscilloscope Calibration Option
5080A/SC/MEG –  Multi-Product Calibrator with Megohm Meter and Oscilloscope Calibration Option
5080/CAL –  5080/CAL calibration software

Knowledge

Product Manuals
application/pdf icon 5080/CAL Users Manual (2.15 MB)
application/pdf icon 5080A Calibration (5.31 MB)
application/pdf icon 5080A Calibration Supplement (82.25 KB)
application/pdf icon 5080A Getting Started Manual (English) (1015.83 KB)
application/pdf icon 5080A Getting Started Supplement (103.43 KB)
application/pdf icon 5080A Operators Manual (English) (8.74 MB)
application/pdf icon 5080A Operators Supplement (103.68 KB)
application/pdf icon 5080A/MEG Megohm Option Users Manual (English) (1.58 MB)
application/pdf icon 5080A/SC Oscilloscope Calibration Option Users Manual (English) (962.17 KB)
Brochures
application/pdf icon 5080A Multiproduct Calibrator Brochure (English) (2.34 MB)
application/pdf icon Electrical Calibration Workload Matrix (581.08 KB)
Application Notes
application/pdf icon Calibrating Analog Meters with the 5080A Multi-Product Calibrator (1.55 MB)
Specifications
application/pdf icon 5080A Multi-Product Calibrator Extended Specifications (554.64 KB)
application/pdf icon 5080A Multi-Product Calibrator MegOhm Meter Calibration Option Extended Specifications (497.76 KB)
application/pdf icon 5080A Multi-Product Calibrator Oscilloscope Calibration Option Extended Specifications (395.92 KB)
Press Releases
New Fluke 5080A Multi-Product Calibrator Helps Ensure Accuracy of Both Analog and Digital Test Instruments