5080A Megohm Meter Calibration Option

ขอใบเสนอราคา

 

 

Highlights

Megohm meter calibration option for the 5080A Multi-Product Calibrator

The 5080A Megohm Meter Calibration Option enables the 5080A Multi-Product Calibrator to calibrate the high voltage output of megohm meters and other portable and bench insulation testers. It sources high ohms, high voltage resistors up to 18 G Ohms. It also measures high voltage outputs.

Operation is easy

You access the 5080A Megohm Meter Calibration Option’s capabilities simply by selecting “MEG” mode on the calibrator’s front panel. All outputs are conveniently located on the calibrator’s output block, to make hook-ups easy and to minimize the errors contributed by cabling.

Specifications

Insulation Resistance Ranges: 10 kohm to 10 Gohm, plus 18 Gohm singe value Best 1-year specification: 0.2%
Insulation Voltage Ranges: 0 V to 1575 V dc peak
Best 1-year specification: 1%
Continuity Resistance Ranges: 1 ohm to 5.9 kohm (16 values)
Best 1-year specification: 0.1%
Continuity Current Ranges: 700 mA max
Best 1-year specification: 1.2%

Models and Accessories

Model Name Description
5080A->5080A/MEG Upgrade 5080A to 5080A/SC
Installable only at Fluke service centers for extra installation and calibration cost.

Knowledge and Information

Product Manuals
5080A Service
application/pdf icon 5080A/MEG Megohm Option Users Manual (English)