5627A Precision Thermometer, RTD Temperature Probe

ขอใบเสนอราคา

 

 

Description

5627A Precision Thermometer, RTD Temperature Probe

One of the best features of this sensor is that it conforms to the standard 385 curve, letting you use your DIN/IEC RTD meters fully. Why use a precision thermometer or RTD temperature probe that’s less accurate than your meter?

The 5627A is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. (Bend, if any, should be specified at time of order.)

Six- and nine-inch 5627A precision thermometers are calibrated at –196 °C, –38 °C, 0 °C, 200 °C, and 300 °C. For 12-inch versions of these precision thermometers, the point at 300 °C is replaced by a calibration point at 420 °C.

Each probe is individually calibrated and includes a NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.

This RTD temperature probe is an excellent value for a precision thermometer. It has the price-to-accuracy and price-to-durability ratios you should demand in all of the precision thermometers you buy!

Highlights

Durable PRTs with temperature range to 420 °C and accuracy to 0.025 °C

  • Vibration and shock resistant
  • NVLAP-accredited calibration included, lab code 200706-0

When buying a PRT or RTD temperature probe, performance isn’t the only criterion you need to look at. The real issues are price-to-accuracy and price-to-durability ratios.

5627A probes have a temperature range up to 420 °C and an accuracy as good as ± 0.025 °C. They come in three different lengths. (Both six- and nine-inch models cover –200 °C to 300 °C.) Each precision thermometer is shipped with its ITS-90 coefficients and a calibration table in 1 °C increments.

Specifications

Specifications

Resistance Nominal 100 Ω
Temperature Coefficient 0.00385 Ω/Ω/ °C nominal
Temperature Range −200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
Drift Rate (k=2) ± 0.04 °C at 0 °C after 100 hours at 420 °C
Sheath Material 316 Stainless Steel
Leads PTFE-insulated, nickel-plated stranded copper, 22 AWG
Termination Specify. See Ordering Information.
Time Constant Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Bending Radius Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.)
Calibration Includes manufacturer’s NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from -196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included.
Immersion At least 100 mm (4 in) recommended
Calibrated Accuracy† (k=2) ± 0.026 °C at -196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C
Size 5627A-12  : 305 mm x 6.35 mm (12 in x 1/4 in)
5627A-9  : 229 mm x 4.7 mm (9 in x 3/16 in)
5627A-6  : 152 mm x 4.7 mm (6 in x 3/16 in)
†Includes calibration uncertainty and 100 hr drift.

Models and Accessories

Model Name Description
5627A-6-X Secondary PRT, 152 mm x 4.7 mm (6 x 3/16 in), –200 °C to 300 °C
(17025 accredited calibration included. Traceable to NIST standards.)
X = termination. Specify “A” (INFO-CON for 914X), “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), “G” (gold pins), “I” (INFO-CON for 1521 or 1522), “J” (banana plugs), “L” (mini spade lugs), “M” (mini banana plugs), “P” (INFO-CON for 1523 or 1524), or “S” (spade lugs).
5627A-9-X Secondary PRT, 229 mm x 4.7 mm (9 x 3/16 in), –200 °C to 300 °C
(17025 accredited calibration included. Traceable to NIST standards.)
X = termination. Specify “A” (INFO-CON for 914X), “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), “G” (gold pins), “I” (INFO-CON for 1521 or 1522), “J” (banana plugs), “L” (mini spade lugs), “M” (mini banana plugs), “P” (INFO-CON for 1523 or 1524), or “S” (spade lugs).
5627A-12-X Secondary PRT, 305 mm x 6.35 mm (12 x 1/4 in), –200 °C to 420 °C
(17025 accredited calibration included. Traceable to NIST standards.)
X = termination. Specify “A” (INFO-CON for 914X), “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), “G” (gold pins), “I” (INFO-CON for 1521 or 1522), “J” (banana plugs), “L” (mini spade lugs), “M” (mini banana plugs), “P” (INFO-CON for 1523 or 1524), or “S” (spade lugs).

Accessories common to all models  :

Accessory Description
2601 Probe Carrying Case, Plastic 
2609 Probe Carrying Case, Plastic (for probes 635 mm [25 in] in length)