Additel 223A
Documenting Process Calibrator
A highly integrated multifunction calibrator featuring several patented technologies, the 223A is an ultra-compact, rugged, and easy to use hand-held device for sourcing, simulating and measuring pressure, temperature, and electrical signals. Its smartphone-like menu and interface make the operation simpler and easier. With HART communication capability, the 223A is ideal for calibrating, maintaining, and troubleshooting HART instrumentations. Automation and documentation capabilities make the 223A a turnkey solution.
The 223A multifunction calibrator is a process tool for measuring, sourcing and simulating mA, mV, V, RTDs, thermocouples, ohms, frequency, pulses and pressure, captures switch values and provides 24 V loop power.
Electrical Specifications
Measurement Accuracy
|
|||||
|
Range
|
Resolution
|
Accuracy
|
||
Voltage DC
|
±75.0000 mV
±30.0000 V
|
0.11µV
0.1mV
|
0.01%RD+3.75 μV
0.01%RD+1.5 mV
|
||
Current DC
|
±30.0000 mA
|
0.1µA
|
0.01%rdg + 1.5 μA
|
||
Resistance
|
Two-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD + 0.02 Ω
|
|
Three-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD + 0.02 Ω
|
||
Four-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.01%RD + 0.02 Ω
|
||
Two-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.02%RD + 0.2 Ω
|
||
Three-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.02%RD + 0.2 Ω
|
||
Four-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.01%RD + 0.2 Ω
|
||
Frequency
|
1 to 50000.0 Hz
|
0.1Hz
|
0.005%RD + 1 Hz
|
||
Pulse
|
0 to 999999
|
1
|
N/A
|
||
Limit Switch
|
If the switch has detective voltage, its range is +3V to +24V
|
||||
Source Accuracy
|
|||||
Voltage DC
|
-10.000 to 75.000mV 0 to 12.0000 V
|
1μV
0.1mV
|
0.02%RD + 4.25 μV
0.02%RD + 0.6 mV |
||
Current DC
|
0 to 22.000 mA
|
1µA
|
0.02%RD + 1.1 μA
|
||
Resistance
|
1 to 400.00 Ω
1 to 4000.0 Ω
|
10mΩ
100mΩ
|
0.02%RD + 0.02 Ω
0.03%RD + 0.4 Ω |
||
Frequency
|
0 to 50000.0 Hz
|
0.1Hz
|
0.005%RD + 1 Hz
|
||
Pulse
|
0 to 999999
|
1
|
N/A
|
||
DC24V
|
N/A
|
N/A
|
0.5V
|
General Specifications
Environmental Specifications
|
|
Operating Temperature
|
-10°C to 50°C
|
Storage Temperature
|
-20°C to 60°C
|
Humidity
|
<90%, non-condensing
|
Safety Specifications
|
|
European Compliance
|
CE Mark
|
Mechanical Specifications
|
|
Display
|
3.5 inch TFT color screen
|
Electrical Connection
|
φ4mm sockets and flat mini-jack thermocouple socket
|
RS232 Interface
|
standard RS232-DB9 socket
|
Size
|
3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm)
|
Weight
|
1.6 lb (0.7 kg)
|
Power Supply
|
Polymer Li-ion rechargeable battery, or 10V DC adaptor
|
Battery
|
Rechargeable Li-ion battery (included)
|
Battery Life
|
15 hours uninterrupted use
Battery life will be reduced when 24V is applied
|
Battery Charge
|
110V/220V external power adapter (included)
|
Pressure Specification
|
|
The 160 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet.
|
Temperature Specifications
Thermocouple Measurement and Source accuracy
|
||||||||||
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
|||||||
Measure
|
Source
|
|||||||||
S
|
IEC 584
|
-50 to 1768
|
-50 to400
|
1.0
|
1.1
|
|||||
400 to 1000
|
0.6
|
0.6
|
||||||||
1000 to1768
|
0.7
|
0.8
|
||||||||
R
|
IEC 584
|
-50 to 1768
|
-50 to 200
|
1.4
|
1.4
|
|||||
200 to 500
|
0.6
|
0.6
|
||||||||
500 to 1768
|
0.6
|
0.7
|
||||||||
B
|
IEC 584
|
0 to 1820
|
50 to 450
|
3.8
|
3.8
|
|||||
450 to 800
|
0.9
|
0.9
|
||||||||
800 to 1820
|
0.6
|
0.7
|
||||||||
K
|
IEC 584
|
-270 to 1372
|
-250 to -200
|
1.0
|
1.1
|
|||||
-200 to -100
|
0.4
|
0.5
|
||||||||
-100 to 600
|
0.3
|
0.3
|
||||||||
600 to 1372
|
0.4
|
0.5
|
||||||||
N
|
IEC 584
|
-270 to 1300
|
-250 to -200
|
1.5
|
1.6
|
|||||
-200 to -100
|
0.5
|
0.6
|
||||||||
-100 to 1300
|
0.4
|
0.5
|
||||||||
E
|
IEC 584
|
-270 to 1000
|
-250 to -200
|
0.6
|
0.7
|
|||||
-200 to -100
|
0.3
|
0.3
|
||||||||
-100 to 0
|
0.2
|
0.2
|
||||||||
0 to 700
|
0.2
|
0.3
|
||||||||
700 to 1000
|
0.2
|
0.4
|
||||||||
J
|
IEC 584
|
-270 to 1200
|
-210 to -100
|
0.3
|
0.3
|
|||||
-100 to 1200
|
0.3
|
0.4
|
||||||||
T
|
IEC 584
|
-270 to 400
|
-250 to -200
|
0.8
|
0.9
|
|||||
-200 to 0
|
0.4
|
0.4
|
||||||||
0 to 400
|
0.2
|
0.2
|
||||||||
C
|
ASTM E988
|
0 to 2315
|
0 to 1000
|
0.5
|
0.5
|
|||||
1000 to 1800
|
0.7
|
0.9
|
||||||||
1800 to 2315
|
1.0
|
1.4
|
||||||||
D
|
ASTM E988
|
0 to 2320
|
0 to 100
|
0.5
|
0.5
|
|||||
100 to 1100
|
0.4
|
0.5
|
||||||||
1100 to 2000
|
0.6
|
0.9
|
||||||||
2000 to 2320
|
0.9
|
1.3
|
||||||||
G
|
ASTM E1751
|
0 to 2315
|
0 to 200
|
2.4
|
2.4
|
|||||
200 to 400
|
0.5
|
0.5
|
||||||||
400 to 1400
|
0.4
|
0.5
|
||||||||
1400 to 2315
|
0.7
|
1.0
|
||||||||
L
|
DIN43710
|
-200 to 900
|
-200 to -100
|
0.2
|
0.3
|
|||||
-100 to 400
|
0.2
|
0.2
|
||||||||
400 to 900
|
0.2
|
0.3
|
||||||||
U
|
DIN43710
|
-200 to 600
|
-200 to 0
|
0.4
|
0.4
|
|||||
0 to 600
|
0.2
|
0.3
|
||||||||
RTD Measurement and Source accuracy
|
||||||||||
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
|||||||
Measure (2W/3W)
|
Measure (4W)
|
Source
|
||||||||
Pt10(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.65
|
0.60
|
0.65
|
||||
200 to 600
|
0.82
|
0.72
|
0.82
|
|||||||
600 to 850
|
0.96
|
0.82
|
0.96
|
|||||||
PT100
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
||||
200 to 600
|
0.26
|
0.16
|
0.26
|
|||||||
600 to 850
|
0.34
|
0.20
|
0.34
|
|||||||
Pt100
(3916)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
||||
200 to 600
|
0.26
|
0.16
|
0.26
|
|||||||
600 to 850
|
0.33
|
0.20
|
0.33
|
|||||||
Pt200(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.37
|
0.32
|
0.69
|
||||
200 to 600
|
0.51
|
0.41
|
0.92
|
|||||||
600 to 850
|
0.61
|
0.48
|
1.08
|
|||||||
Pt500(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.20
|
0.16
|
0.36
|
||||
200 to 600
|
0.32
|
0.22
|
0.54
|
|||||||
600 to 850
|
0.40
|
0.27
|
0.67
|
|||||||
Pt1000
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.1
|
0.05
|
0.26
|
||||
200 to 600
|
0.2
|
0.10
|
0.42
|
|||||||
600 to 850
|
0.27
|
0.14
|
0.54
|
|||||||
Cu100(428)
|
GOST6651-94
|
-50 to 150
|
-50 to150
|
0.12
|
0.09
|
0.12
|
||||
Cu50(428)
|
GOST6651-94
|
-50 to 150
|
-50 to 150
|
0.17
|
0.13
|
0.17
|
||||
Cu10(427)
|
Minco Application Aid #18
|
-100 to 260
|
-100 to 260
|
0.61
|
0.56
|
0.61
|
||||
Ni120(672)
|
Edison curve #7
|
-100 to 260
|
-100 to 260
|
0.07
|
0.05
|
0.07
|
||||
Ni100(618)
|
DIN 43760
|
-100 to 260
|
-100 to 260
|
0.08
|
0.06
|
0.08
|
110V/220V external power adapter
|
1 pc
|
Chargeable Li-ion battery
|
1 pc
|
Test leads
|
2 sets (4 pcs)
|
short circuit cable
|
1 sets (2 pcs)
|
Manual
|
1 pc
|
NIST traceable calibration certificate
|
1 pc
|
Model number
|
Description
|
ADT160A |
Intelligent Digital Pressure Modules
|
9060
|
Pressure module connection cable
|
9050
|
USB to RS232 (DB-9 Male) Adapter
|
9080
|
Cold Junction compensation kits (including TC plug, compensation cable, S,R,B,K,J,T,E,N)
|
9712
|
Spare chargeable Li-ion battery for multifunction calibrator
|
9816
|
110V/220V external power adapter for multifunction calibrator
|
9906
|
Carrying case for multifunction calibrator
|
9510
|
Additel/Cal Task management software for multifunction calibrator
|