Additel 780
For years, we’ve provided the most durable, accurate, quality pressure calibration products for field applications. The Additel 780 controller incorporates the same durability, accuracy, and quality into a new bench top controller packed with features and functionality that is remarkably easy to use. The Additel 780 offers two base ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the maximum controlling range of the controller. Each configuration includes a control sensor in which you can preselect the sensor range best suited your application. External and internal sensors can be used which allows for expanded range and accuracy capability in the future.
Modular Design
Each unit comes with one Intelligent Pressure Module configured to the many range offerings provided. Standard accuracy sensors (ADT160A-CPXXX) are silicon pressure sensors with a 1 year accuracy of 0.02% FS. The precision quartz based sensors improve the 1 year accuracy specification to 0.005% of reading + 0.005% FS. Each sensor has been specially aged, tested and screened before assembly. After assembly each sensor is temperature compensated over the range of -10°C to 50°C. The Additel 780 allows for one internal pressure sensor and one external pressure sensor. The modular design of this unit provides for interchangeability of both the internal and external sensors with other Intelligent Pressure Modules.In addition to the Intelligent Pressure Modules, the Additel 780 has a built in barometric sensor. This allows for switching between gauge pressures to absolute pressures.
Standalone Solution
Typical pressure controllers will require a nitrogen bottle for the gas supply which make it difficult to move the controller around without having to move or connect to another bottle. The Additel 780 is unique in that with the optional electric pump, you can generate pressures to 1,000 psi (70 bar) without the need of a gas bottle. Add the booster to the configuration and 3,000 psi (200 bar) is achievable with the Additel 780 controller.
Process Functionality
The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.
The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.
Measurement Specifications
Model Number | ADT780-1K | ADT780-3K |
Pressure range | -14.5 to 1,000 psi (-0.95 to 70 bar) |
-14.5 to 3,000 psi (-0.95 to 200 bar)[1] |
Control stability | 0.003% FS | |
Precision (includes 1 year stability) |
See pressure range table | |
1 Year stability | 0.01% FS standard accuracy sensors 0.005% FS precision accuracy sensors | |
Media | clean Gas | |
Over-range indication | 103% to 120% (based on sensor) | |
Resolution | 4, 5, 6, or 7 digits (user selectable)[2] | |
Pressure type | Gauge, Absolute | |
Warm up time | 15 minutes | |
Typical pressure settling time (within 0.003%FS) |
20 seconds (10%FS pressure change assuming 50ml test volume) | |
Measurement units | Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2, kg/cm2, mtorr, torr, atm, lb/ft2, tsi, user selectable |
|
Minimum control pressure[3] | 0.0001 psi | 0.001 psi |
[1] Gas booster or HP gas supply required to reach 3,000 psi (200 bar). Booster ordered separatel.
[2] Digit resolution for precision model only.
[3] Dependent on pressure module.
Barometric Measurement Specifications
Sensor/Mode | Gauge | Absolute |
ADT160A-02-CPXX | N/A | 40Pa[1] |
ADT160A-01-APXXQ | 3Pa[2] | N/A |
Barometer range (60~110)KPa, the accuracy is 40Pa (11Pa is optional)
[1] 40Pa uncertainty (k=2) includes calibration uncertainty, linearity and long term stability (< 30Pa per year). Barometer range of 60 to 120KPa.
[2] Combined linearity, hysteresis and repeatability. Add 3Pa when used in gauge mode. When using the ADT160A-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.
Electrical Measure Specifications
Specification | Range | Resolution | Accuracy |
Volts DC | -30 to 30 V | 0.1 mV | ±0.01% rdg + 1.5 mV |
-300 to 300 mV | 1 uV | ±0.01% rdg + 15 uV | |
Current DC | -30 to 30 mA | 0.1 uA | ±0.01% rdg + 1.5 uA |
Switch test | If the switch has detected voltage, the range is from 3 – 24 V | ||
DC 24V output | 24V ± 0.24V, max 30 mA |
Physical Specifications
Specification | ADT780-1K & ADT780-3K |
Power | 100 – 240V, 50/60 Hz |
Pressure ports | G1/8 F |
Storage temperature | -20°C to 70°C |
Operating environment | 0-95% RH non-condensing |
Display | 7 inch (17.8 cm) color, touch screen display |
Weight | 33 lbs (15 kg) |
Dimensions(DWH) | 16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm) |
Mounting | Standard desktop, optional rack mount |
Shock | 4G |
Vibration | 1G 10Hz~500Hz |
Other Specifications
WiFi specifications | 802.11 b, g, and n |
Stability indicator | User selectable |
Typical slew rate | Typically 20 seconds |
Operating modes | Control, measure, and vent |
Display modes | Controller – show pressure indication and control Calibrator – shows pressure indication, control, and electrical measurement |
Localization | English, Chinese (simplified), German, Spanish, French, Italian, Portuguese, Russian, Japanese |
Conformity | CE |
Calibration certification | NIST-traceable certificate of calibration with data included |
Electric Pump and Booster
ADT780-1K-EPUMP Electric Pump
Generate Vacuum to 1,000 psi (70 bar)
Built-in filter and liquid trap
Pressure range | -13.5 to 1,080 psi (-0.95 to 75 bar) |
Weight | 69.6 lbs (31.6 kg) |
Size(WHD) | 17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm) |
Power | 100 to 240V |
Media | Air |
Volume | 375 mL |
Outlet port | G 1/8F |
Storage temperature | -20°C to 70°C |
Operating environment | 0 – 90% RH non-condensing |
Mounting | Standard desktop, optional rack mount |
Control operation | 2 LED displays with pressure limit settings |
Typical max pressure time | Approx 2 min with ADT780 Controller |
ADT780-3K-BST Booster
3,000 psi (200 bar) Gas booster specification
Minimum inlet pressure | 300 psi (20 bar) |
Out pressure range | 300 to 3,600 psi (250 bar) |
Inlet flow rate | 500ml/min |
Size (WHD) | 17.8 X 7.3 X 23.9 inch (451 X 185 X 607 mm) |
Volume | 375 mL |
Weight | 74 lbs (33.6 kg) |
Power | 100 to 240V |
Supply media | Nitrogen or clean gas |
The Range of ADT160A Intelligent Digital Pressure Modules
P/N | Pressure Range | Media | Accuracy | Pressure Rating | Suggested Controller Compatibility | Internal/ External Use | ||
psia | bar.a | Burst | Over Pressure | |||||
Precision Quartz Sensors[1] |
||||||||
AP45Q | 0 to 45 | 3.0 | G | 0.005% rdg + 0.005% FS | 3x | 1.2x | ADT780-1K only | Both |
AP100Q | 0 to 100 | 7.0 | G | 0.005% rdg + 0.005% FS | 3x | 1.2x | ADT780-1K only | Both |
AP400Q | 0 to 400 | 28 | G | 0.005% rdg + 0.005% FS | 3x | 1.2x | Both | Both |
AP500Q | 0 to 500 | 35 | G | 0.005% rdg + 0.005% FS | 3x | 1.03x | Both | Both |
AP1KQ | 0 to 1,000 | 70 | G | 0.005% rdg + 0.005% FS | 3x | 1.1x | Both | Both |
AP2KQ | 0 to 2,000 | 140 | G | 0.005% rdg + 0.005% FS | 3x | 1.1x | ADT780-3K only | Both |
AP3KQ | 0 to 3,000 | 200 | G | 0.005% rdg + 0.005% FS | 3x | 1.1x | ADT780-3K only | Both |
Standard Pressure Sensors |
||||||||
P/N | Pressure Range | Media | Accuracy | Pressure Rating | Suggested Controller Compatibility | Internal/ External Use | ||
psig | bar.g | Burst | Over Pressure | |||||
CP10 | ±10 | ±0.7 | G | 0.02% FS | 3x | 1.2x | ADT780-1K only | Both |
CP15 | ±15 | ±1.0 | G | 0.02% FS | 3x | 1.2x | ADT780-1K only | Both |
CP30 | -15 to 30 | -1 to 2.0 | G | 0.02% FS | 3x | 1.2x | ADT780-1K only | Both |
CP50 | -15 to 50 | -1 to 3.5 | G | 0.02% FS | 3x | 1.2x | ADT780-1K only | Both |
CP100 | -15 to 100 | -1 to 7.0 | G,L | 0.02% FS | 3x | 1.2x | ADT780-1K only | Both |
CP300 | -15 to 300 | -1 to 20 | G,L | 0.02% FS | 3x | 1.2x | Both | Both |
CP500 | -15 to 500 | -1 to 35 | G,L | 0.02% FS | 3x | 1.2x | Both | Both |
CP600 | -15 to 600 | -1 to 40 | G,L | 0.02% FS | 3x | 1.2x | Both | Both |
CP1K | -15 to 1,000 | -1 to 70 | G,L | 0.02% FS | 3x | 1.2x | Both | Both |
CP2K | -15 to 2000 | -1 to 140 | G,L | 0.02% FS | 3x | 1.2x | ADT780-3K only | Both |
CP3K | -15 to 3,000 | -1 to 200 | G,L | 0.02% FS | 3x | 1.2x | ADT780-3K only | Both |
CP5K | -15 to 5000 | -1 to 350 | G,L | 0.02% FS | 3x | 1.2x | N/A | External |
CP10K | -15 to 10,000 | -1 to 700 | G,L | 0.02% FS | 2x | 1.2x | N/A | External |
[1] Contact Additel for other range options.
**Low pressure sensors (ADT160A-XX-DPX) available for low pressure and differential pressure measurement. Also available for low pressure control (gauge mode only). Ranges from ±1 inH20 (2.5 mbar) to 300 inH2O (700 mbar).