Temperature Range
|
1950: -200 °C to 500 °C
1960: -200 °C to 670 °C
|
Rtpw
|
Nominal 25 Ω
|
Resistance Ratio
|
W(Ga)>=1.11807
W(Hg)<=0.844235
|
Drift at 0.01°C*
|
1950 <0.002 °C after 100 hours at 500 °C,
<0.004 °C/year typical
1960 <0.003 °C after 100 hours at 661 °C,
<0.005 °C/year typical
|
Repeatability
|
±0.001 °C
|
Thermal Shock
|
±0.001 °C after 10 times thermal cycles from minimum to maximum temperatures
|
Self-heating
|
0.0015 °C at 1 mA current
|
Measurement Current
|
1 mA
|
Sensor Length
|
42 mm
|
Insulation Resistance
|
>1000 MΩ at room temperature
|
Sheath Material
|
Quartz
|
Dimension
|
1950: 7 mm (OD) X 480 mm (L)
1960: 7 mm (OD) X 500 mm (L)
|
External Leads
|
Insulated copper wire, 4 leads, 2.5 meters
|
Termination
|
Gold-plated Spade
|
Handle Dimension
|
21 mm (OD) X 80 mm (L)
|
Optional Calibration
|
NVLAP/ISO17025 accredited calibration and data available per request
|
*Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
Model
|
Description
|
9002
|
Wooden Carrying Case included
|