Fluke 719Pro-300G Electric Pressure Calibrator

ขอใบเสนอราคา

 

 

Description

Fluke 719Pro-300G Electric Pressure Calibrator

  • Unique integrated electric pump for one handed pressure calibration
  • Test pressure switches easily with easy-to-use switch test function
  • Precision pressure adjust vernier for easy and accurate pressure calibration
  • Ideal accuracy and performance (0.025 %) for reference class transmitter calibration
  • 30 PSI, 150 PSI and 300 PSI pressure ranges
  • Programmable pressure calibrator pump limit:
    • Eliminates over-pressurization
    • Allows the user to program target pressures for pumping
  • Extended pressure measurement range with any of 29 700Pxx Pressure Modules
  • Pump cleanout ports reduce pump failures from moisture and fluid damage
  • Fluid trap included with the unit to mitigate chance of fluid damage in at-risk applications
  • Precision temperature measurement with optional RTD probe

Highlights

The 719Pro includes a full functioning loop calibrator that sources, simulates and measures mA signals and more making it the ideal test tool for calibrating high accuracy transmitters, pressure switches and pressure gauges.

Get the ultimate in measurement flexibility with the large backlit screen, which displays three parameters at once:

  • Pressure measurement from internal or external sensor
  • Sourced/simulated or measured mA values
  • Temperature measured by optional RTD probe

Specifications

Specifications
Pressure Measurement
Range -12 to 300 PSI / -0.8 to 20 bar
Resolution 0.01 PSI, 0.001 bar
Accuracy (1 year) ± 0.025 % of full scale, 6 months ± 0.035 % of full scale, 
1 year Dry Air and non-corrosive gas only
Temperature effect (all ranges) Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C 
No effect on accuracy on all functions from 15 °C to 35 °C

Function

Range

Resolution

Accuracy (1 year)

Comment

mA dc (simulation and measurement) 0-24 mA dc 0.001 mA ± 0.015 % of reading ± 2 counts 1000 ohm maximum load in mA source, 26 V dc maximum voltage in mA simulate
Volts dc (measurement only) 30 V dc 0.001 V ± 0.015 % of reading ± 2 counts No AC voltage measurement, do not exceed 30 V dc
Temperature (measurement only, Pt100 RTD) -50 to 150 °C (-50 °F to 302 °F) 0.01 °C 0.01 °F Temperature ± 0.1 °C (0.2 °F) ± 0.25 °C (± 0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory)
Loop power supply 24 V dc N/A 24 mA at 24 V
Temperature effect (all functions)     Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C No effect on accuracy on all functions from 15 °C to 35 °C
Environmental
Operating temperature -10 °C to +50 °C
Storage temperature (with batteries) Per battery manufacturer’s specification within storage specification below
Storage temperature (without batteries) -20 °C to +60 °C
Power Requirements
Battery Eight (8) standard Alkaline AA batteries
Battery life 300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 PSI; 100 pump cycles to 300 PSI minimum
Physical
Dimensions (HxWxD) 24.13 cm x 11.18 cm x 7 cm (9.5 in x 4.4 in x 2.8 in)
Weight With holster 1.2 kg (2.5 lb)
Connections
Pressure 1/8 inch female NPT
Electrical Standard banana jacks
RTD 4 pin miniature circular
External module 6 pin miniature circular
Engineering Units (Configurable)
psi, bar, mbar, kPa, kgcm2, cmH2O@4 °C, cmH2O@20 °C, mH2O@4 °C, mH2O@20 °C,
inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, mmHg@0 °C, inHg@0 °C, ftH2O@4 °C, ftH2O@20 °C, ftH2O@60 °F

Models and Accessories

Model Name Description
Fluke-719Pro-300G Electric Pressure Calibrator, 300 PSIG
Fluke-719Pro-30G Electric Pressure Calibrator, 30 PSIG
Fluke-719Pro-150G Electric Pressure Calibrator, 150 PSIG

Accessories common to all models

Accessory Description
Fluke 720RTD RTD Probe for 721 and 719Pro
Fluke-71XTrap Liquid and Dirt Trap for 718 and 719Pro
750Pxx  Pressure Module Series

Knowledge and Information

Brochures

Data Sheets