HC2-HK25 / HC2-HK40

ขอใบเสนอราคา

 

 

 

OVERVIEW

The handheld probe is especially suitable for mobile measurements of high temperatures. It measures humidity and temperature and calculates the dew/frost point.

Applications
Climatic and temperature chambers, dryers, air ducts

Features
• Range of application: -100…200 °C / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
• Adjusted at 23 °C and 10, 35, 80 %rh

TECHNICAL DATA

PROBE TYPE High-Temperature Handheld Probe
OPERATING RANGE -40…100 °C / 0…100 %rh
ACCURACY AT 23 ±5 °C ±0.8 %rh / ±0.1 K
ACCURACY W. ADJUSTMENT PROFILE “STANDARD” at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
POWER SUPPLY / POWER CONSUMPTION 3.2…5 VDC, Calibrated @ 3.3 VDC / ~4.5 mA
LONG-TERM STABILITY <1 %rh / year
HUMIDITY SENSOR Hygromer IN-1
TEMPERATURE SENSOR PT100 1/3 DIN Class B
PROBE PROTECTION Filter carrier, brass nickel plated
FILTER CARTRIDGE Wire mesh, 20…25 um
RESPONSE TIME T 63 <15 sec.
ALLOWED WIND SPEED 20 m/s, with filter
PSYCHROMETRIC CALCULATIONS Dew or frost point
REPRODUCIBILITY AIRCHIP <0.02 %rh / 0.01 K
EXPLOITATION SOFTWARE ROTRONIC HW4
PROBE ADJUSTMENT BY SOFTWARE 1 Point & multipoint %rh & °C
FIRMWARE UPDATE With internet and HW4 software
USER INFORMATION from Software HW4 (devicemanager)
DATA PROCESSING VIA HW4 with interface cables
DEVICE PROTECTION BY PASSWORD Yes
TYPE OF OUTPUT SIGNALS 2 x 0…1 VDC
ANALOGUE OUTPUTS SCALEABLE BY THE USER Yes
ANALOG OUTPUT SIGNAL (STANDARD) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
DIGITAL OUTPUT SIGNAL (STANDARD) UART
SERVICE INTERFACE UART
REVERSE VOLTAGE PROTECTION Mechanically in the connector
MAX. CABLE LENGTH 2 m
PROBE MATERIAL PEEK
ENCLOSURE MATERIAL POM handle / sensor tube: PPS, black
FDA/GAMP COMPATIBILITY FDA 21 CFR Part 11 and GAMP compatible
WEIGHT 210 g / 240 g

DOWNLOADS

HC2 – Data sheet (345.94 KB)6/7/2016

HC2 – Manual v1.38 (777.23 KB)5/11/2016

HC2 – Short instruction manual (459.52 KB)9/23/2015