HC2-P05

ขอใบเสนอราคา

 

 

 

Category: Tag:

OVERVIEW

The insertion probe is suitable for measurement in dust-free bulk materials, bricks, concrete, etc. It measures humidity and temperature and calculates the dew/frost point.

Applications

Water activity measurements, a portable handheld instruments and data logger

Features

  • Range of application: -40…85 °C / 0…100 %RH
  • Digital interface (UART) and scalable analog outputs, 0…1 V
  • Standard configuration: 0…1 V = -40…60 °C / 0…100 %RH
  • Adjusted at 23 °C and 10, 35, 80 %RH

TECHNICAL DATA

FOR USE WITH HygroLab C1, HP23-AW-A
OPERATING RANGE -40…85 °C / 0…100 %rh
RANGE OF APPLICATION -40…85 °C (-40…185 °F) / 0…100 %RH
ACCURACY AT 23 ±5 °C ± 0.015 aw / 1.5 %rh / ±0.3 K
POWER SUPPLY 3.2…5 VDC, calibrated at 3.3 VDC, current: ~4.5 mA
LONG-TERM STABILITY <1 %rh / year
HUMIDITY SENSOR HYGROMER® WA-1 / Pt-100, 1/3 DIN Class B
TEMPERATURE DEPENDENCE 0.001 K/K, 0.0002 aw/K
FILTER TYPE No filter available
RESPONSE TIME T 63 <15 sec.
PSYCHROMETRIC CALCULATIONS Dew or frost point
START-UP TIME 1,5 s
MEASUREMENT INTERVAL 1 s
MATERIAL Stainless steel DIN 1.4305 (probe), POM (handle)
IP PROTECTION RATING IP30
FDA / GAMP FDA 21 CFR part 11 / GAMP5
CE / EMC CONFORM 2007 / 108 / EG
DIMENSIONS ∅ 5 x 200 mm
WEIGHT 160 g

DOWNLOADS

HC2 – Data sheet (345.94 KB)6/7/2016

HC2 – Manual v1.38 (777.23 KB)5/11/2016

HC2-P05 CE-EMV (149.48 KB)11/13/2017

Water activity – Data sheet (1010.63 KB)12/13/2017