HC2-S3H

ขอใบเสนอราคา

 

Category: Tag:

OVERVIEW

The HC2-SH / HC2-S3H fulfills the highest demands on measuring accuracy. It measures humidity and temperature and calculates the dew/frost point.

Applications
HVAC, food industry, building services equipment, paper, textile and pharmaceutical industries

Features
• Accuracy: ±0.5 %rh, ±0.1 K, at 23 °C ±5 K
• Range of application: -50…100 °C / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
• ISO17025:2005 / ANSI/NCSL Z540-1-1994 Part 1 accredited by NVLAP Lab Code 201016-0 calibration certificate. Humidity calibration performed at 5%, 15%, 25% 35%, 50% 65%, 80%RH@23ºC and 95%RH@20ºC. Temperature calibration performed at 23ºC.

APPLICATIONS

  • Meteorological

TECHNICAL DATA

PROBE TYPE Meteorological probe
OPERATING RANGE -50…100 °C / 0…100 %rh
ACCURACY AT 23 ±5 °C ±0.5 %rh / ±0.1 K
ACCURACY W. ADJUSTMENT PROFILE “HIGH PRECISION” at 23 °C / 10, 20, 30, 40, 50, 60, 70, 80, 90 %rh ± 0.5 %rh / 0.1 K
POWER SUPPLY / POWER CONSUMPTION 3.2…5 VDC, Calibration @ 3.3 VDC / ~4.5 mA
LONG-TERM STABILITY <1 %rh / year
HUMIDITY SENSOR Hygromer IN-1
PROBE PROTECTION Polycarbonate plastic cage, white
FILTER CARTRIDGE Polyethylene dust filter, white, 40 um
RESPONSE TIME T 63 <20 sec.
ALLOWED WIND SPEED 20 m/s, with filter
PSYCHROMETRIC CALCULATIONS Dew or frost point
REPRODUCIBILITY AIRCHIP <0.02 %rh / 0.01 K
EXPLOITATION SOFTWARE ROTRONIC HW4
PROBE ADJUSTMENT BY SOFTWARE 1 Point & multipoint %rh & °C
FIRMWARE UPDATE With internet and HW4 software
USER INFORMATION from Software HW4 (devicemanager)
DATA PROCESSING VIA HW4 with interface cables
DEVICE PROTECTION BY PASSWORD Yes
TYPE OF OUTPUT SIGNALS 2 x 0…1 VDC
ANALOGUE OUTPUTS SCALEABLE BY THE USER Yes
ANALOG OUTPUT SIGNAL (STANDARD) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
DIGITAL OUTPUT SIGNAL (STANDARD) UART
COMMUNICATION INTERFACES UART
SERVICE INTERFACE UART
REVERSE VOLTAGE PROTECTION Mechanically in the connector
MAX. CABLE LENGTH Up to 5 m with passive cable. Up to 100 m with active cable.
ENCLOSURE MATERIAL Polycarbonate
FDA/GAMP COMPATIBILITY FDA 21 CFR Part 11 and GAMP compatible
DIMENSIONS ø 15 x 83 mm
WEIGHT 10 g

DOWNLOADS

Campbell Scientific – Customer Application (Meteo) (103.38 KB)4/3/2017

HC2 – Data sheet (345.94 KB)6/7/2016

HC2 – Manual v1.38 (777.23 KB)5/11/2016

HC2 – Short instruction manual (459.52 KB)9/23/2015

HC2-S / HC2-S3 – CE declaration (164.13 KB)5/13/2016